With the CL515-PLUS you can
check and calibrate all your RTD
instruments and measure RTD
Sensors. Automatic indication of
connections on the display for
simple hookups.
Stop carrying around a decade box
and RTD resistance tables. The
CL515-PLUS works with the RTDs
you use including Platinum 100
(alpha = 3850, 3902, 3926) & 1000
(alpha = 3850, 3750) Ohm, Copper
10 & 50 Ohm, Nickel 100 and 120
Ohm. Easily set any value quickly to
within 0.1° with the adjustable digital
potentiometer “DIAL” plus store any
three temperatures for instant recall
with the check switch.
Calibrate quickly with automatic
output stepping, choose between
2, 3, 5, 11 and 21 steps to
automatically increment the output
in 100%, 50%, 25%, 10% or 5% of
span. Select the step time to match
your system from 5, 6, 7, 8, 10, 15,
20, 25, 30 and 60 seconds.
Carry it without worry—it comes
protected with a rubber boot and
rugged, low profile switch. Easy to
operate even in the dark areas of
the plant with the backlit display.
SPECIFICATIONS (Unless otherwise indicated all specifications are rated from a nominal 23°C, 70% RH for 1 year from calibration) Accuracy: ±(0.015% of Reading + 0.05 Ω) Temperature Drift: ± 0.01% of span outside of 23°C ±10°C (73°C ±18°F) Operating Temperature Range: -25 to 60°C (-10 to 140°F) Relative Humidity Range: 10% ≤RH ≤90% (0 to 35°C), non-condensing, 10% ≤RH ≤70% (35 to 60°C), non-condensing Dimensions: 143 L x 76 W x 41 mmH (5.63 x 3.00 x 1.60") Weight: 0.34 kg (12.1 oz) including boot and batteries Power: 4 "AA" Alkaline 1.5V (LR6) (included) Battery Life: 50 Hours low Battery low battery indication with nominal 1 hour of operation left. Protection against misconnection over-voltage protection to 60 Vdc (rated for 30 seconds) Display: High contrast graphic liquid crystal display. LED backlighting for use in low lit areas.
READ Excitation: Current 0.5 mA nominal Normal Mode Rejection: 50/60 Hz, 50 dB Common Mode Rejection: 50/60 Hz, 120 dB SOURCE
Accuracy: From 1 to 10.2 mA external excitation Current ± (0.015% of Reading + 0.05 Ω) Below: 1 mA of external excitation Current: ± (0.015% of reading + 0.025 mV/mA excitation current + 0.05 Ω) Resistance Ranges: 0.00 to 410.00, 410.1 to 4001.0 Ω Allowable Excitation Current Range
<410 Ohms: 10.2 mA max; steady or pulsed/intermittent 410 to 4001 Ohms: 1 mA max; steady or pulsed/intermittent Pulsed Excitation: Current compatibility DC to 0.01 second pulse width 0.025 mV
† 所有金额以RMB显示 注: Comes complete with rubber boot, 4 “AA” alkaline batteries, NIST certificate of calibration, carrying case, RTD wire kit, 2 red and 2 black leads with retractable shield banana plugs and spade lugs and operator’s manual. 订购示例: (1) CL515-PLUS Universal RTD Calibrator, ¥17,955.89