omega-qr-code 微 信 公 众 号
咨询热线 400-820-0559

Calibrator for Thermocouple and RTDs (PT100)

CL543B

RTD Calibrator | Thermocouple Calibrator | CL543B

RTD Calibrator | Thermocouple Calibrator

¥16,932.33 CL543B

CE
  • Calibrate Thermocouple Instruments to 0.1 and 0.01, °F and °C
    Types J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), (T-DIN) and P (Platinel): -13.000 to 80.000 mV
  • Calibrate RTD Instruments to 0.1 and 0.01, °F and °C
    Pt 10, 50, 100, 200, 500 and 1000 (3850)
    Pt 100 (alpha = 3902, 3916, 3926)
    Cu 10 and 50 and Ni 120
  • Calibrate with Confidence
    Resolution to 0.01, °F and °C; Accurate to 0.2°
    Cold Junction Sensor Accurate to 0.05°C
    NIST Traceable Cold Junction Sensor in Isothermal Block
  • Troubleshoot RTD Sensors and Instruments
    Automatic Detection of 2-, 3- or 4-Wire Connections Indicates When Individual Wire Connections are Open
    Verify Proper Instrument Operation by Measuring the Excitation Current
  • Fast Easy Calibrations
    Instantly Select 3 Outputs with the EZ-CHECK Switch
    Easily Find Trip Points with 2 Speed EZ-DIAL Knob and Automatic Stepping
  • Verify Resistance, Cold Junction Temperature and Output
    Secondary Display of Resistance for RTDs and Compensated millivolt Setting, and Cold Junction Temperature for Thermocouple
  • Use in the Plant, Field or Shop
    Includes Rubber Boot and mV Test Leads
    Low Profile Switches Resist Accidental Damage
    Brass Screws for Thermocouple Wires Plus Miniature Thermocouple Socket
  • Perform Heat Treating Surveys and Accuracy Tests
    Meets or Exceeds the Requirement of AMS 2750 as Both a Secondary Standard Instrument and as a Field Test Instrument
  • NIST Cert (No Points)
精密校准设备和冰点参考 - 查看相关产品

产品描述

THERMOCOUPLE CONNECTIONS
Simulating or reading thermocouples requires the use of thermocouple or extension grade thermocouple wire. Plug thermocouple wires into the miniature thermocouple jack or place bare thermocouple wires onto the brass block under the screws.

The CL543B has two banana jacks (1+ and 2-) mounted in the top end of the housing. These are not temperature compensated and are to be used only for millivolt signals.

RTD CONNECTIONS
Simulating or reading RTDs uses copper wire. Plug 2-, 3- or 4-wires into the corresponding jacks on the calibrator. For RTD source the CL543B simulates the (+) RTD from jacks 1 and 4 and the (-) RTD from jacks 2 and 3. When reading an RTD sensor the CL543B uses patented circuitry to automatically detects if 2-, 3- or 4-wires are connected. This is helpful to troubleshoot sensor connections.

Specifications
Operating Temperature Range: -20 to 60°C (-5 to 140°F)
Storage Temperature Range: -30 to 60°C (-22 to 140°F)
Temperature Effect: ≤±25 ppm/°C of range
Cold Junction Sensor: ≤±0.01% °/° from ambient temperature (°C or °F)
Relative Humidity Range:
    10% ≤RH ≤90% (0 to 35°C), non-condensing
    10% ≤RH ≤70% (35 to 60°C), non-condensing
Normal Mode Rejection: 50/60 Hz, 50 dB
Common Mode Rejection: 50/60 Hz, 120 dB
Dimensions: 143 L x 76 W x 41 mm H (5.63 x 3 x 1.60")
Weight: 0.34 kg (12.1 oz) with boot and batteries
Batteries: Four “AA” alkaline 1.5V (LR6) (included), optional NiMh rechargeable battery kit [120 Vac (North America only); charger, 4 NiMh batteries, AC and DC cords
Battery Life: 50 hours, low battery indication with nominal 1 hour of operation left
Over-Voltage Protection: Up to 60 vrms (rated for 30 seconds), protection against misconnection
Alerts: Red LED displays OVERLOAD or out of range conditions
Display: High contrast graphic liquid crystal display, LED backlighting for use in low lit areas
Voltage Source
Ranges and Resolution: -13.000 to 80.000 mV and -13.0000 to 80.0000 mV
Accuracy: ±(0.008% of setting + 0.006 mV)
Source Current: ≥10 mA
Output Impedance: <0.3Ω
RMS Noise: ≤±0.0005 mV from 0.1 to 10 Hz
Short Circuit Duration: Infinite Voltage Read
Ranges and Resolution: -13.000 to 80.000 mV and -13.0000 to 80.0000 mV
Accuracy: ±(0.008% of reading + 0.006 mV)
Input resistance: ≥10 MΩ Thermocouple Source
Accuracy: ±(0.008% of setting + 0.006 mV)
Cold Junction Compensation: ±0.05°C (±0.09°F); thermistor traceable to NIST for 11 years
Output Impedance: <0.3Ω
Source Current: >10 mA (drives 80 mV into 10Ω)
RMS Noise: ≤±0.0005 mV from 0.1 to 10 Hz
Thermocouple Read
Accuracy: ±(0.008% of reading + 0.006 mV)
Cold Junction Compensation: ±0.09°F (±0.05°C); thermistor traceable to NIST for 11 years
Input Impedance: >10 MΩ
Open Thermocouple Threshold: 10,000Ω nominal
Pulse: <10 microamp pulse for 300 milliseconds
RTD and O Source
3- and 4-Wire Accuracy:
    From 1 to 10.2 mA External Excitation Current: ±(0.015% of setting + 0.05Ω)
    Below 1 mA of External Excitation Current: Add (0.025 mV/mA excitation current)
2-Wire Accuracy: Add 0.1Ω to 3-wire and 4-wire accuracy
Resistance Ranges:
    400Ω Range: 0.00 to 401.00 and 0.000 to 401.000
    4000Ω Range: 0.0 to 4010.0 and 0.00 to 4010.00
RMS Noise:
    400Ω Range: ≤±0.005Ω from 0.1 to 10 Hz
    4000Ω Range: ≤±0.05Ω from 0.1 to 10 Hz
Allowable Excitation Current Range:
    400Ω Range: 10.2 mA max; steady or pulsed/intermittent
    4000Ω Range: 1 mA max; steady or pulsed/intermittent
Pulsed Excitation Current Compatibility: DC to 0.01 second pulse width, 0.025 mV, mA excitation current
RTD and Ohms Read
Resistance Ranges:
    400Ω Range: 0.00 to 401.00 and 0.000 to 401.000
    4000Ω Range: 0.0 to 4010.0 and 0.00 to 4010.00
Accuracy: ±(0.015% of reading + 0.05Ω)
Excitation Current: 0.9 mA to 401Ω, 0.4 mA to 4010Ω (nominal)
订购 所有价格均为人民币税前价格,不包括13%增值税和运费!(指定型号)
网上下单立享5%折扣!
产品编号产品描述
数量
产品描述
Quick Ship
CL543B
¥16,932.33
货期: 2 周
 
Thermocouple and RTD temperature calibrator
所有金额以RMB显示
注: Comes complete with NIST Cert (no points), 4 “AA” alkaline batteries, rubber boot, deluxe hands free carrying case,mA/V test leads, 1 red and 1 black lead with banana plugs and alligator clips, RTD wire kit, 2 red and 2 black leads with banana plugs and spade lugs, and operator’s manual.