Handheld Temperature Calibrator for all Thermocouple Types |
CL20 Series |
|
|
|
|
|
|
|
|
The CL23/24/25/26/27 Models has been discontinued. Please see the CL940A/ CL945A/ CL543B as a possible alternative or contact Sales
|
|
0.5°F Accuracy Over the Range of 18°C to 28°C (64°F to 82°F)
|
|
NIST Traceable Certificate of Calibration
|
|
Simulates and Measures Thermocouples, RTDs, Thermistors and Ohms
|
|
0.1°F/0.1°C Resolution
|
|
10 Memory Locations for Saving Meter Readings and/or Calibrator Outputs
|
|
Step Calibrator, Meter Hold, Thermometer Functions
|
|
Ramp Function (Models CL25 and CL27)
|
|
Review PDF File (Link Below) for Full Specifications & Features
|
View related products - Precision Calibration Equipment and Ice Point References
|
| |
|
|
|
The CL23/24/25/26/27 Models has been discontinued. Please see the CL940A/ CL945A/ CL543B as a possible alternative or contact Sales
The OMEGA CL20 series calibrator/
thermometers simulate, measure and
record RTD, ohm, thermocouple and
thermistor signals–all in one meter.
They outperform the competition in the
lab and on the factory floor with 0.3°C
(0.5°F) accuracy and in ambient
temperatures from 18 to 28°C
(64 to 82°F).
This powerful handheld meter allows
the user to calibrate in the field as well
as in the lab and offers many additional
functions that translate into costeffectiveness,
increased productivity
and long-lasting versatility and
reliability.
The 5-digit LCD display indicates all
active functions and allows users to
have 0.1° resolution in both Celsius and
Fahrenheit. This convenient display
takes the guesswork out of which
sensor has been selected, displaying
°C or °F and operating mode.
Programming a CL20 series calibrator
is a snap. The keypad allows direct
entry of calibrator setpoints. Setpoints
can also be entered from an outside
source by inputting temperatures or
voltages through the thermometer
mode. There are no reprogramming
hassles. The program won’t be lost
when the meter is shut off. All values
are retained and can be retrieved for
fast repeat use.
These calibrators feature 10 memory
locations that can instantly recall any
programmed value, and the step
function lets the user run through the
test points at the push of a button. This
efficient feature saves time by allowing
the user to simply enter the output
value and calibrate.
The ramp function of Models CL25 and
CL27 makes calibrating switch and
alarm setpoints quick and easy. By
entering the upper limit, lower limit and
step size, the meter automatically
ramps between the limits. There is also
a manual ramp function.
No matter what thermocouple type is
used, one of these meters can handle
it. Features to choose from include 100
ohm and 1000 ohm RTD and 2252 ohm
thermistor capabilities.
|
| |
Need to comply with ISO-9000 reporting specifications? These meters can help! All
conform to the temperature/voltage tables of the National Institute of Standards and Technology
(N.I.S.T.). OMEGA provides a free NIST Traceable Certificate of Calibration for your records.
Since ISO-9000 requires that calibration be done at the equipment site, these meters are a
perfect choice. With 0.3°C (0.5F°) accuracy in an ambient temperature range of 18 to 28°C
(64 to 82°F), you’re not confined to the lab. Take it everywhere. The ABS plastic case protects it from dust and dirt.
|
|
Specifications
Display: 5-digit LCD custom indicators for calibration type, units
hold, recall, store, operating mode, bal, loaded memory locations
(up to 10)
Keypad: Momentary switches with tactile feedback, select
power, thermocouple type, units, resolution, hold, store/recall,
stop/clear, operating mode
Configuration Retention: selected operating mode,
thermocouple type, units, resolution and memory contents
retained on power off
Diagnostics: low battery, open input, invalid keypad entry,
out of range, internal hardware fault
Accuracy: see ordering matrix for details
Ambient Temperature: 0 to 50°C (32 to 122°F), 0 to 90% RH
Reading Rate: 1 per second
Input Current: 50 nA typical(meter mode)
Output Load: 100 ohms, min. (calibrator mode)
Max. Common Mode Voltage: 42 V peak to earth ground
Power: 9 Vdc battery, alkaline, included
Battery Life: 16 hrs alkaline
Dimensions: 178 H x 74 W x 28 mm D (7.0 x 2.9 x 1.1")
Weight: 340 g (12 oz)
|
|
|
CL20 Series Temperature Measurement and Calibration Range
Input Type | Range | Resolution |
K | 200 to 1371°C | -328 to 2502°F | 0.1° |
J | -210 to 760°C | -346 to 1400°F | 0.1° |
T | -200 to 399°C | -328 to 752°F | 0.1° |
E | -230 to 999°C | -382 to 1832°F | 0.1° |
N | 200 to 1299°C | -328 to 2372°F | 0.1° |
B | 499 to 1815°C | 932 to 3308°F | 0.1° |
R | 0 to 1768°C | 32 to 3214°F | 0.1° |
S | 0 to 1768°C | 32 to 3214°F | 0.1° |
G | 299 to 2315°C | 572 to 4201°F | 0.1° |
C | 0 to 2315°C | 32 to 4201°F | 0.1° |
D | 0 to 2315°C | 32 to 4201°F | 0.1° |
RTD | -200 to 849°C | -328 to 1562°F | 0.1° |
Series 400 Thermistor 2252Ohm | -40 to 150°C | -40 to 302°F | 0.1° |
CL26 and CL27 Resistance Measurement and Calibration Outputs
Ohms Range | Resolution | Models |
0-100Ω | 0.01Ω | CL26 & CL27 |
0-1000Ω | 0.1Ω | CL27 |
0-100,000Ω | 1.0Ω | CL26 |
|
Model Number | Features |
TC Types | Platinum RTDs | Thermistor | Resistance 0 to: | Resistance Resolution | Accuracy Calibrator & Thermometer | Ramp Feature |
CL23A | K, J, T | --- | --- | --- | --- | ±0.3°C from -46 to 677°C 0.04% over 677°C ±0.5°C under -46°C | No |
CL24 | K, J, T, E | --- | --- | --- | --- | 0.04% over 677°C ±0.5°C under -46°C | No |
CL25 | K,J,T,E,N,R,S,B,G,C,D | --- | --- | --- | --- | ±0.3°C full scale K,J,T,E,N 0.8°C full scale B,R,S,G,C,D | Yes |
CL26 | K, J, T ,E | 100Ω | 400 Series 2252Ω | 100 kΩ 100Ω | 1Ω 0.01Ω | ±0.3°C from -46 to 677°C 0.04% over 677°C ±0.5°C under -46°C Ωs: 0.04% of range ±0.3°C RTD & thermistor | No |
CL27 | K,J,T,E,N,R,S,B,G,C,D | 100Ω 1000Ω | --- | 1kΩ 100Ω | 0.1Ω 0.01Ω | ±0.3°C full scale K,J,T,E,N 0.8°C full scale B,R,S,G,C,D Ωs: 0.02% of range 0.1°C full scale RTD | Yes |
|
|
|
|
|
|
† Price are shown in RMB.
Note: Each CL20 handheld meter is supplied as a complete kit with 4 beaded wire thermocouples (J, K, T, or E), 4 TAS standard to mini thermocouple adaptors (J,K,T,E), 4 calibration cables (J, K, T, or E), shock resistant holster, belt clip, integral tilt stand/handle, wrist strap, rugged carrying case, and certificate of calibration. The CL26 and CL27 handheld meter also includes one 3-wire RTD calibration cable, one 2-wire resistance cable and one TA4F RTD adaptor. The CL23A handheld meter does not include type E thermocouple, adaptor or cable. To order additional spare thermocouple calibration cables of different type, change "K" in part number to desired type, e.g. CL-300-CABLE-T-2.
|
|
|
|
|
| |
|
Back to top
|
|
|