Integrated Circuit IC Temperature Probe |
OM-2628 |
|
|
|
|
Integrated Circuit Temperature Probe
|
Click for larger image. |
|
|
¥
1,100.00
OM-2628-C1
|
|
High Precalibrated Accuracy: ±0.5°C (0.9°F) Maximum @ 25°C
|
|
Excellent Linearity: 0.15°C (0.27°F) Maximum [0 to 70°C (32 to 158°F)]
|
|
Wide Operating Temperature Range: -25 to 105°C (-13 to 221°F)
|
|
Single Supply Operation: +4 to +30V
|
|
Excellent Repeatability and Stability
|
|
High Level Output: 1 µA/K (278.2µA at 25°C)
|
|
Two Terminal Monolithic IC Provides Current Output Proportional to Temperature
|
|
Minimal Self-Heating Errors
|
|
38 and 153 mm (1.5 and 6") Standard Lengths
|
|
Typical Appilications Include: Appliance Temperature Sensing Automotive Temperature Measurement and Control HVAC System Monitoring Industrial Temperature Control
|
View related products - Thermistor Elements, Probes and Assemblies
|
| |
|
|
|
The OM-2628 IC probe features
the integrated circuit transducer
AD592CN. This IC solid state sensor is a two
terminal monolithic integrated
circuit temperature transducer
that provides an output current
proportional to absolute
temperature. For a wide range
of supply voltages the transducer
acts as a high impedance
temperature dependent current
source of 1 µA/K. Improved design
and laser wafer trimming of the
IC’s thin film resistors allows the
AD592CN solid state sensor to achieve absolute
accuracy levels and nonlinearity
errors previously unattainable at
a comparable price.
The OM-2628 probe can
be employed in applications
between -25 and 105°C
(-13 to 221°F) where conventional
temperature sensors (i.e.,
thermistor, RTD, thermocouple,
diode) are currently being used.
Expensive linearization circuitry,
precision voltage references,
bridge components, resistance
measuring circuitry and cold
junction compensation are not
required with the OM-2628 probe.
The OM-2628 solid state sensor is particularly useful
in remote sensing applications,
the AD592CN is immune to
voltage drops and voltage noise
over long lines due to its high
impedance current output.
Specifications
Accuracy
Calibration Error @ 25°C (77°F) TA = 0 to 70°C (32 to 158°F): 0.3°C (0.54°F) typical, 0.5°C (0.9°F) maximum
Error Over Temperature: 0.4°C (0.72°F) typical, 0.8°C (1.4°F) maximum
Nonlinearity TA = -25 to 105°C (-13 to 221°F): 0.05°C (0.09°F) typical, 0.15°C (0.27°F) maximum
Error Over Temperature: 0.5°C (0.9°F) typical, 1°C (1.8°F) maximum
Nonlinearity: 0.1°C (0.18°F) typical, 0.35°C (0.63°F) maximum
Output Characteristics
Nominal Current Output @ 25°C (298.2 K): 298.2 µA typical
Temperature Coefficient: 1 µA/°C typical
Repeatability: 0.1°C (0.18°F) typical
Long Term Stability: 0.1°C/month (0.18°F) typical
Absolute Maximum Ratings
Operating Temperature: -25°C (-13°F) minimum, 105°C (221°F) maximum
Package Temperature: -45°C (-49°F) minimum, 125°C (257°F) maximum
Forward Voltage (+ to -): 44V maximum
Reverse Voltage (- to +) Lead Temperature: 20V maximum
Soldering, 10 Seconds: 572°C (1062°F) maximum
Power Supply
Operating Voltage Range Power
Supply Rejection: 4V minimum, 30V maximum
4V < Vs < 5V: 0.5°C (0.9°F)/V maximum
5V < Vs < 15V: 0.2°C (0.36°F)/V maximum
15V < Vs < 30V: 0.1°C (0.18°F)/V maximum
|
|
|
|
|
|
† Price are shown in RMB.
Ordering Example: (1) OM-2628-C1 38 L x 6.4 mm diameter (1.5 x 1/4") AD592CN temperature probe with 0.91 m (36") leads, ¥1,100.00
|
|
|
|
|
| |
Related Links -
Related Products
|
|
|
Related Products
|
|
DP41:
|