Pressure Transmitter, Rangeable Wet/Wet Differential |
PX5200 Rangeable dp Transmitter |
|
|
|
|
|
|
|
|
Compact Size
|
|
NEMA 4X (IP65) Environmental Rating
|
|
Backlit LCD for Easy Reading
|
|
Display Can Be Scaled to Any Units
|
|
Display Rotates in 90° Increments to Fit Location
|
|
Stainless Steel, FKM, and Ceramic Wetted Parts
|
|
Flow measurement Totalization/Square Root Extraction Function
|
|
Quick Ranging with Internal Push-Buttons
|
|
Scale to Display in Any Engineering Units
|
|
MIN/MAX and HOLD Functions
|
|
Loop Check Function to Check Circuit
|
|
Program Lock Function
|
View related products - Pressure Transducers
|
| |
|
|
|
This item is being discontinued.
The OMEGA™ PX5200 is a
compact wet/wet differential pressure
dP transmitter, ideal for flow and
tank level applications where
reliable, low dP measurements are
required. Equipped with a solid state
sensor and a 316 Stainless Steel
isolating diaphragm. The PX5200
can monitor a wide variety of wet or
dry media. Available in unidirectional
ranges from 0 - 4 to 0 - 400 inH2O
and bi-directional ranges up to ±200 inH2O, the PX5200 operates
on line (static) pressures up
to 300 psi.
A rugged aluminum NEMA 4X
(IP65) enclosure measures only
65 mm (2.56') in diameter and
incorporates a backlit LCD display
for local reading.
The output is a two-wire
configurable 4 to 20 mA signal that
can be adjusted to re-range the
transmitter or "tare" unwanted fluid
head preloads. In addition, this
instrument offers flow measurement/square root extraction, where the
indicated flow rate and analog signal
can be displayed. Rangeability for
zero and span are from -10 to
110% FS URL (Upper Range Limit).
This means a 20 inH2O transducer
can have the zero set between -2
inH2O to FS 22 inH2O. Span can
be set over that range also. For
example zero could be set for 2
inH2O and 22 inH2O maximum so
the output would be 4 mA at
2 inH2O, midpoint 12 mA would be
12 inH2O and 20 mA would be
22 inH2O.
SPECIFICATIONS
Output: 4 to 20 mAdc (2-wire)
Supply Voltage: 12 to 32 Vdc Vmin=12V+(0.022Ax loop res)
Insulation Resistance: 50 Vdc (>100 MΩ)
Rangeability/Adjustments:
Zero: -10% to 110% FS
Span: -10% to 110% FS
Accuracy: ±0.5% FS URL (upper range limit) @23°C includes linearity, hysteresis and repeatability
Note: Accuracy and output resolution based upon FS URL (upper range limit) value.
Stability: ±0.25% FS/year
Response Time: 100 ms (adjustable)
Output Resolution: 0.1% FS URL
Display: 4-digit, 10 mm LCD with LED backlight
Display Accuracy: ±0.5% FS URL (upper range limit) + 1 digit
Storage Temperature: -15 to 65°C (5 to 150°F)
Operating Temperature: -10 to 60°C (14 to 140°F)
Compensated Temperature: -10 to 60°C (14 to 140°F)
Temperature Effects (Over Comp Range): ±0.03% FS URL /°C (from 23°C reference)
Static (Line) Pressure (All Ranges): 300 psi proof; 800 psi burst
Static (Line) Pressure Effects:
≥20 inH2O, ±8 inH2O: ±0.03% FS/ 100 psi
8 inH2O, ±4 inH2O: ±0.07% FS/ 100 psi
4 inH2O: ±0.15% FS/100 psi
Single Side (Differential) Limits:
≤8 inH2O, ±4 inH2O: 30 psid proof; 130 psi burst
≥20 inH2O, ±8 inH2O: 100 psid proof; 130 psi burst
Vibration: 5 g's 150 Hz
Shock: 10 g's 16 ms
Pressure Connection: ¼ NPT female
Electrical Termination: ½ NPT female conduit fitting
Environmental Rating: NEMA 4X (IP65)
Media: Fluids and gasses compatible with 316SS, FKM, and alumina ceramic
Weight: Approx 454 g (1 lb)
|
|
|
|
|
|
† Price are shown in RMB.
Note: Comes complete with mounting bracket and operator's manual.
|
|
|
|
|
| |
Related Links -
Manual, Related Products
|
|
|
Product Manual
|
Download
PX5200 Series - Pressure Transmitter (Long Version Manual)
|
Download
PX5200 Series - Pressure Transmitter (Quick Start Manual)
|
Related Products
|
|
|
|
|
|
|