DIGITAL PRESSURE GAUGES METRIC FITTINGS AND RANGES |
DPGM409DIFF Series |
|
|
|
|
DIGITAL PRESSURE GAUGES with METRIC FITTINGS AND RANGES
|
Click for larger image. |
|
|
¥
13,305.00
DPGM409-010BDWU
|
|
Wet/Wet Differential Pressure Measurement
|
|
5-Point NIST Traceable Calibration
|
|
Easy to Read 25 mm (1") Backlit LCD Display
|
|
High 0.08% Accuracy
|
|
Hi and Lo Alarms Standard
|
|
MIN/MAX/AVG Reading
|
|
Analog Output: User selectable 0 to 5 Vdc, 0 to 10 Vdc or 4 to 20 mA (external 24 Vdc power supply required for all analog output types)
|
|
Bar Graph on Display
|
|
Free Downloadable Software
|
|
Wireless Transmitter Option Sends Reading to Remote Location
|
|
CE Certified
|
|
Battery Powered with Long Battery Life
|
|
NEMA 4X (IP65) Enclosure
|
|
Metric Ranges and Pressure Connections
|
View related products - Pressure Gauges
|
| |
|
|
|
OMEGA’s DPGM409 digital
differential pressure gauge
incorporates a rugged 316 SS
enclosure designed for washdown
and marine applications with a
precision 0.08% accuracy pressure
transducer and 316L SS wetted
parts and body. The result is an
extremely durable digital pressure
gauge. You get all the durability
and stability of the micromachined
silicon sensor in a rugged NEMA 4X
(IP65) package.
Wireless Option: Add suffix "-W"to part number for extra cost option
NOTE: Models with wireless option are approved for use in the U.S., Canada and Europe only.
SPECIFICATIONS
CE Compliant: Meets EN1326-1: 2006
for industrial locations with analog
output cables <3 m (9.8') and supplied
ferrite core
Accuracy (Combined Linearity,
Hysteresis and Repeatability): ±0.08% BSL
Storage Temperature: -40 to 82°C (-40 to 180°F)
Operating Temperature: -18 to 66°C (0 to 150°F)
Thermal Effects (Over Operating Range)
Zero Balance:
Ranges 25 mb to 70 mb: ±1% span
Ranges >70 mb: ±0.5% span
Span Setting:
Ranges 25 mb to 70 mb: ±1% span
Ranges >70 mb: ±0.5% span
Minimum Resistance Between Transducer Body and Any Wire:
100M Ω @ 50 Vdc
Pressure Cycles: 1 million minimum
Long Term Stability (1-yr.): ±0.1% FS typical
Analog Output: User selectable
0 to 5 Vdc, 0 to 10 Vdc or 4 to 20 mA
(external 24 Vdc power supply required
for all analog output types)
Notes: 1. Bidirectional ranges have a
unidirectional analog output equal to their full
span; i.e. -15 to +15 psi = 0 to 10 Vdc.
2. Analog output is independently scalable
Set jumper to Select Current or Voltage:
Software selection of 0 to 5 Vdc, 0 to
10 Vdc, or 4 to 20 mA
Analog Output Cable: 2.7 m (9') with
ferrite core (included)
Analog Output Accuracy:
0.08% typical (0.15% maximum in RF
field @ 10V/m)
Line/Static Pressure:
500 psi maximum applied to both
sides simultaneously
Proof Pressure (Differential)
25 mb: = 10 times range
70 mb: = 6 times range
170 mb to 50 Bar: = 4 times range
70 Bar Range: = 3 times range
Hi-Side Containment Pressure (Differential):
25 to 350 mb: To 1000 psi
1 to 70 bar: To 3000 psi
Display: 4-digit LCD with user
selectable backlighting, (9999 count)
25 mm (1") character height
Sample/Display Range: User
selectable from 0.38 to 30 seconds
(preset to 1/second)
Computer Interface: USB connection
for set-up, cable included
Power: One 3.6V lithium 8.4 Ah
capacity (C cell) included (two included
with wireless option) (replacement
battery model no. BATT-C-3.6V)
Battery Life (Typical): Up to 4-years
Enclosure Material: 316 Stainless
Steel and ABS center gasket
Enclosure Finish: Electropolished
Enclosure Environmental Rating:
Weatherproof, NEMA 4X (IP65)
Sensor Wetted Parts: 316L stainless steel
Pressure Connection: G ¼ male
Dimensions
Display: 115 D x 89 W x 74 mm H (4.5 x 3.5 x 2.9")
Sensor: 72 to 89 mm L (2.8 to 3.5")
below display, 80 mm (3.1") wide"
Weight: 1 kg (2.2 lb) typical depending upon configuration
Front Panel Functions
Backlight: Manually turns backlight on/off
Min/Max: Recalls min/max values
Zero/Clear: Zeros display or clears min/max
Software Settings (via USB)
(USB connection is for set-up only)
Units: psi, inHg, inH2O, bar, mbar
Lock: Allows the lock-out of front
panel functions
Alarms: User selectable “High” and
“Low” alarm limits (open collector)
Backlight: On/Off, 10 seconds,
30 seconds, 1 minute, 5 minute
Wireless Transmitter (Optional):
Channel number, transmission rate,
alarms, sensor offset, chart recording,
data logging
Calibrate: Zero and Span
Overrange Indication: “_ _ _ _”
Wireless Option Specifications
Transmit Sample Rate: User
programmable from 1 sample/2 min to
1 sample/2 sec
Radio Frequency (RF) Transceiver
Carrier: ISM 2.4 GHz
RF Output: 10 dBm (10 mW)
Range of RF Link: Up to 120 m (400')
outdoors (line of sight); Up to 40 m
(130') indoors/urban
Included Software: Requires
Windows® 7 (32 bit); XP, Vista
Data Transmission to Host: Pressure
reading, ambient temperature reading,
RF transmit strength and battery level
|
|
|
|
|
|
† Price are shown in RMB.
Note: Comes complete with operator’s manual, battery (2 with wireless option), downloadable set-up and configuration software, USB cable (with wireless option) charting and databasing software with common drivers, USB cable, 2.7 m (8.9') analog output cable with ferrite core and 5-point NIST traceable calibration certificate. To order wireless option add suffix “-W” to model number for additional cost.
Ordering Example: (1) DPGM409-010BDWU 10 bar wet/wet differential pressure gauge, ¥13,305.00,
|
|
|
|
|
| |
Related Links -
Related Products
|
|
|
Related Products
|
|
DPGM409S Series:
|
DIGITAL PRESSURE GAUGE for SANITARY/CLEAN-IN-PLACE APPLICATIONS HIGH 0.08% ACCURACY
|
|
|
|
|
|
DPG409 Series:
|
Digital Pressure Gauge (DPG) with Analog Output and Optional Wireless Transmitter
|
|
|
|
|
|
|
|
|